TDS5000B Online Programmer Guide
-- 5 4 0 --
MEASUrement:IMMed:REFLevel:ABSolute:MID2
Description
This command sets or queries the mid reference level for the ”to” wa-
veform when taking a delay measurement, and is the 50% reference
level when MEASUrement:IMMed:REFLevel:METHod is set to Abso-
lute. Note that this command affects the results of delay measurements.
Group
Measurement
Related Commands
MEASUrement:IMMed:REFLevel:METHod
Syntax 1
MEASUrement:IMMed:REFLevel:ABSolute:MID2 <NR3>
Syntax 2
MEASUrement:IMMed:REFLevel:ABSolute:MID2?
Arguments
H <NR3>
This is the mid2 reference level, in volts. The default is 0.0 V.
Example 1
MEASUREMENT:IMMED:REFLEVEL:ABSOLUTE:MID2 0.5
This command sets the mid reference level for the delay waveform to
0.5 V.
Example 2
MEASUREMENT:IMMED:REFLEVEL:ABSOLUTE:MID2?
This query might return
:MEASUREMENT:IMMED:REFLEVEL:ABSOLUTE:MID2
0.5000E+00
indicating that the absolute mid2 reference level is set to 0.5 V.