Page #755 background image
TDS5000B Online Programmer Guide
-- 7 3 5 --
TRIGger:A:PULse:TRANsition:THReshold:HIGH
Description
This command sets or queries the upper (most positive) transition trig-
ger threshold. This command is equivalent to selecting Transition Setup
from the Trig menu and then setting the desired Upper Level voltage.
Group
Trigger
Related Commands
TRIGger:A:PULse:TRANsition:THReshold:LOW
Syntax 1
TRIGger:A:PULse:TRANsition:THReshold:HIGH
<NR3>
Syntax 2
TRIGger:A:PULse:TRANsition:THReshold:HIGH?
Argument
H <NR3>
This specifies the upper threshold, in volts.
Example 1
TRIGger:A:PULse:TRANsition:THReshold:HIGH?
This query might return
:TRIGGER:A:PULSE:TRANSITION:THRESHOLD:HIGH
2.0000E+00
indicating that the upper limit of the pulse transition trigger is 2 V.
Example 2
TRIGger:A:PULse:TRANsition:THReshold:HIGH
120E-3
This command sets the upper limit of the pulse transition trigger to
120 mV.