EasyManuals Logo

Tektronix TDS5000B Series Online Programmer Guide

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #388 background imageLoading...
Page #388 background image
TDS5000B Online Programmer Guide
-- 3 6 8 --
HORizontal[:MAIn]:SCAle
Description
This command sets the time per division for the time base or returns its
horizontal scale on the display and is identical to the HORizon-
tal:SCAle command. The specified scale value is rounded to a valid
scale setting. This command is equivalent to selecting Position/Scale
from the Horiz/Acq menu and then choosing a Scale value.
Group
Horizontal
Related Commands
HORizontal:SCAle
Syntax 1
HORizontal:MAIn:SCAle <NR3>
Syntax 2
HORizontal:MAIn:SCAle?
Argument
H <NR3>
This is the time per division. The range is from 200 ps through 40 s.
Example 1
HORizontal:MAIn:SCAle 2E-6
This command sets the main scale to 2µs per division.
Example 2
HORizontal:MAIn:SCAle?
This query might return:
:HORIZONTAL:MAIN:SCALE 2.0000E-06
This indicates that the main scale is currently set to 2 µs per division.