Page #772 background imageLoading...
Page #772 background image
TDS5000B Online Programmer Guide
-- 7 5 2 --
TRIGger:A:PULse:WINdow:THReshold:LOW
Description
This command sets or queries the lower limit for the window trigger.
This command is equivalent to selecting Window Setup from the Trig
menu and setting the Lower Level voltage.
Group
Trigger
Syntax 1
TRIGger:A:PULse:WINdow:THReshold:LOW <NR3>
Syntax 2
TRIGger:A:PULse:WINdow:THReshold:LOW?
Argument
H <NR3>
This specifies the threshold value in volts.
Example 1
TRIGger:A:PULse:WINdow:THReshold:LOW?
This query might return
:TRIGGER:A:PULSE:WINDOW:THRESHOLD:LOW
1.2000E-01
indicating that the lower limit of the window trigger is currently set to
120 mV.
Example 2
TRIGger:A:PULse:WINdow:THReshold:LOW 0.8
This command sets the lower threshold of the window trigger to 0.8 V.