EasyManua.ls Logo

Tektronix TDS5000B Series

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TDS5000B Online Programmer Guide
-- 7 5 2 --
TRIGger:A:PULse:WINdow:THReshold:LOW
Description
This command sets or queries the lower limit for the window trigger.
This command is equivalent to selecting Window Setup from the Trig
menu and setting the Lower Level voltage.
Group
Trigger
Syntax 1
TRIGger:A:PULse:WINdow:THReshold:LOW <NR3>
Syntax 2
TRIGger:A:PULse:WINdow:THReshold:LOW?
Argument
H <NR3>
This specifies the threshold value in volts.
Example 1
TRIGger:A:PULse:WINdow:THReshold:LOW?
This query might return
:TRIGGER:A:PULSE:WINDOW:THRESHOLD:LOW
1.2000E-01
indicating that the lower limit of the window trigger is currently set to
120 mV.
Example 2
TRIGger:A:PULse:WINdow:THReshold:LOW 0.8
This command sets the lower threshold of the window trigger to 0.8 V.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Related product manuals