-- 8 1 0 --
WFMInpre:YOFf
Description
This command sets or queries the vertical position of the incoming wa-
veform in digitizing levels. Variations in this number are analogous to
changing the vertical position of the waveform. For those formats in
which WFMInpre:BYT_Nr is important (all nonfloating point formats),
this command must take the location of the binary point implied by
BYT_NR into consideration.
Group
Waveform Transfer
Related Commands
WFMInpre:BYT_Nr, WFMInpre:YMUlt, WFMOutpre:YOFf
Syntax 1
WFMInpre:YOFf <NR3>
Syntax 2
WFMInpre:YOFf?
Argument
H <NR3>
This is the vertical offset in digitizing levels.
Example 1
WFMInpre:YOFf?
This query might return
:WFMINPRE:YOFF 25
indicating the vertical position of the incoming waveform in digitizing
levels.
Example 2
WFMInpre:YOFf 50
This command specifies that the zero reference point for the incoming
waveform is 50 digitizing levels (2 divisions) above the center of the
data range.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals