TDS5000B Online Programmer Guide
-- 3 5 7 --
HORizontal:FASTframe:XZEro:FRAme:REF<x>?
Description
This query--only command returns xzero for the specified frame. The
reference waveform is specified by x. The value of x can range from 1
through 4 for four-channel instruments or 1 through 2 for two-channel
instruments.
Xzero is the subsample time between the trigger sample (designated by
WFMOutpre:PT_Off?) and the occurrence of the actual trigger for the
waveform specified.
Group
Horizontal
Related Commands
WFMOutputpre:XZEro?, WFMOutpre:PT_Off?,
HORizontal:FASTframe:XZEro:ALL:CH<x>?,
HORizontal:FASTframe:XZEro:ALL:REF<x>?,
HORizontal:FASTframe:XZEro:FRAme:CH<x>?,
HORizontal:FASTframe:XZEro:REF?,
HORizontal:FASTframe:XZEro:SELECTED:CH<x>?,
HORizontal:FASTframe:XZEro:SELECTED:REF<x>?
Syntax
HORizontal:FASTframe:XZEro:FRAme:REF<x>? <NR1>
Argument
<NR1>
The frame number for which the xzero will be returned on the speci-
fied waveform.
Example
HORizontal:FASTframe:XZEro:FRAme:REF1? 2
This query might return:
:HORIZONTAL:FASTFRAME:XZERO:FRAME:REF1
2,1.4647E-9\n’
This is xzero for the reference 1 waveform, frame 2.