TDS5000B Online Programmer Guide
-- 1 0 4 --
CAL?
Description
This query--only command returns the status of instrument calibration,
including internal signal path calibration, factory calibration, and probe
calibration. This query also returns the time until notification when the
next factory calibration is due. Refer to the service manual for factory
calibration command descriptions.
Group
Calibration
Related Commands
CALibrate:RESults:SPC?
Syntax
CAL?
Example
CAL?
This query might return
:CALIBRATE:INTERNAL:STATUS PASS;:CALI-
BRATE:TEMPERATURE 32;FACTORY:FANCONTROL 2;STA-
TUS PASS;STEPSTATUS FAIL;STEPSTIMULUS
”0.0,0,0,0,END”;NOTIFY:HOURS 10500;YEARS
5.5000;DUE DUE;:CALIBRATE:RESULTS:SPC
PASS;FACTORY PASS;FACTORY:VOLTAGE PASS;FRE-
QUENCY PASS;TRIGGER PASS;:CALIBRATE:PROBES-
TATE:CH1 1;CH2 1;CH3 1;CH4 1