TDS5000B Online Programmer Guide
-- 5 4 5 --
MEASUrement:IMMed:REFLevel:PERCent:MID2
Description
This command sets or queries the percentage (where 100% is equal to
HIGH) used to calculate the mid reference level for the second wave-
form specified when MEASUrement:IMMed:REFLevel:METHod is set
to Percent. Note that this command affects the results of delay measure-
ments.
Group
Measurement
Related Commands
MEASUrement:IMMed:REFLevel:METHod
Syntax 1
MEASUrement:IMMed:REFLevel:PERCent:MID2 <NR1>
Syntax 2
MEASUrement:IMMed:REFLevel:PERCent:MID2?
Arguments
H <NR1>
This is the mid reference level, ranging from 0 to 100%. The default
mid reference level is 50%.
Example 1
MEASUREMENT:IMMED:REFLEVEL:PERCENT:MID2 40
This command sets the mid2 reference level to 40% of HIGH.
Example 2
MEASUREMENT:IMMED:REFLEVEL:PERCENT:MID2?
This query might return
:MEASUREMENT:IMMED:REFLEVEL:PERCENT:MID2 45
Indicating that the percentage mid2 reference level is set to 45% of
HIGH.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Tektronix TDS5000B Series Specifications

General IconGeneral
Channels2 or 4
Vertical Resolution8 bits
Display TypeColor LCD
Vertical Sensitivity1 mV/div to 10 V/div
Input CouplingAC, DC, GND
Timebase Accuracy± 50 ppm
Operating Temperature0 °C to +50 °C
Bandwidthup to 1 GHz
Sample Rateup to 5 GS/s
Display Size10.4 inches
Input Impedance1 MΩ ± 1%
Trigger TypesEdge, Pulse Width, Logic, Runt, Timeout, Setup/Hold
Maximum Input Voltage300 V (DC + peak AC)
InterfaceUSB, GPIB, Ethernet