TDS5000B Online Programmer Guide
-- 1 2 5 --
CH<x>:PRObe:GAIN?
Description
This query--only command returns the gain factor of the probe that is
attached to the specified channel. The channel is specified by x. The
value of x can range from 1 through 4 for four--channel instruments or
1 through 2 for two --channel instruments. The ”gain” of a probe is the
output divided by the input transfer ratio. For example, a common 10x
probe has a gain of 0.1.
Group
Vertical
Related Commands
CH<x>:SCAle
Syntax
CH<x>:PRObe:GAIN?
Example
CH2:PRObe:GAIN?
This query might return
:CH2:PROBE:GAIN 0.1000E+00
indicating that the attached 10x probe delivers 0.1 V to the channel 2
BNC for every 10 V applied to the probe input.