EasyManua.ls Logo

Tektronix TDS5000B Series

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TDS5000B Online Programmer Guide
-- 1 0 8 --
CALibrate:INTERNal:STATus?
Description
This query--only command returns the current status of the instrument
internal signal path calibration for the last SPC operation.
Group
Calibration
Related Commands
*CAL?
Syntax
CALibrate:INTERNal:STATus?
Returns
This query will return one of the following:
H INIT
Indicatesthe instrument has not had internal signal path calibration
run.
H WARMUP
Indicates the instrument has not warmed up for the required time to
perform signal path calibration.
H PASS
Indicates that the signal path calibration completed successfully.
H FAIL
Indicates that the signal path calibration did not complete successful-
ly.
H TEMPDRIFT
Indicates that the instrument has not reached a state of stable tempera-
ture adequate for performing signal path calibration.
H RUNNING
Indicates that the signal path calibration is currently running.
Example
CALibrate:INTERNal:STATus?
This query might return
:CALIBRATE:INTERNAL:STATUS INIT
indicating that the current status of the internal signal path calibration is
that it has not been run.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Related product manuals