TDS5000B Online Programmer Guide
-- 4 4 6 --
MASK:MASKPRE:VPOS
Description
This command sets or returns the nominal vertical position, in divi-
sions, used to vertically position the input channels. The query form of
this command returns the vertical position value of the displayed mask.
The set form of this command only affects the current (displayed)
mask.
Group
Mask
Related Commands
MASK:MASKPRE:VOFFSet, MASK:MASKPRE:VSCAle
Syntax 1
MASK:MASKPRE:VPOS <NR3>
Syntax 2
MASK:MASKPRE:VPOS?
Argument
H <NR3>
Is a floating point number that sets the nominal vertical position in
divisions.
Example
MASK:MASKPRE:VPOS -0.8
Sets the current mask nominal vertical position to -- 0.8 divisions.