TDS5000B Online Programmer Guide
-- 6 9 9 --
TRIGger:A:LOGIc:SETHold:SETTime
Description
This command sets or queries the setup time for setup and hold viola-
tion triggering. This command is equivalent to selecting Setup/Hold
Setup from the Trig menu and then setting the desired Setup Time.
Group
Trigger
Related Commands
TRIGger:A:LOGIc:SETHold:HOLDTime
Syntax 1
TRIGger:A:LOGIc:SETHold:SETTime <NR3>
Syntax 2
TRIGger:A:LOGIc:SETHold:SETTime?
Argument
H <NR3>
This specifies the setup time for setup and hold violation triggering.
Example 1
TRIGger:A:LOGIc:SETHold:SETTime?
This query might return
:TRIGGER:A:LOGIC:SETHOLD:SETTIME 2.0000E-09
indicating that the current setup time for setup and hold trigger is 2 ns.
Example 2
TRIGger:A:LOGIc:SETHold:SETTime 3.0E-6
This command specifies that the current setup time for setup and hold
trigger is 3 µs.