EasyManuals Logo
Home>Tektronix>Test Equipment>TDS5000B Series

Tektronix TDS5000B Series Online Programmer Guide

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #409 background imageLoading...
Page #409 background image
TDS5000B Online Programmer Guide
-- 3 8 9 --
LIMit:LOG
Description
This command sets or queries whether a log file is saved when the
source waveform data exceeds the test limits.
Group
Limit Test
Related Commands
Syntax 1
LIMit:LOG {OFF|ON|NR1}
Syntax 2
LIMit:LOG?
Arguments
H OFF
This argument disables saving a log file when the source waveform
exceeds the test limits.
H ON
This argument enables saving a log file when the source waveform
exceeds the test limits.
H NR1
A 0 disables the save log file feature so that when the source wave-
form data exceeds the limits set by the limit test, a log file is not
saved; any other value enables the save log file feature so that when
the source waveform data exceeds the limits set by the limit test, a log
file is saved. For queries, a 0 is returned if the save log file feature is
off; a 1 is returned if the save log file feature is on.
Example 1
LIMit:LOG 1
This command enables the save lof file feature so that when the source
waveform data exceeds the limits set by the limit test, a lof file is saved.
Example 2
LIMit:LOG?
This query might return:
:LIMIT:LOG 0
This indicates that the save log file feature is disabled and that a log file
will not be saved when the source waveform exceeds the limits of the
limit test.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals