EasyManuals Logo

Tektronix TDS5000B Series Online Programmer Guide

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #424 background imageLoading...
Page #424 background image
TDS5000B Online Programmer Guide
-- 4 0 4 --
MASK:AUTOSet:AUTOAdjust
Description
This command sets or returns whether the instrument settings are auto-
matically matched to signal characteristics and specific mask require-
ments. It controls what happens at the end of a mask autoset. If ON, the
horz/vert auto adjustment is run ONCE to potentially improve the result
of mask autoset.
This command, like all the MASK:AUTOSET commands, affects only
an autoset on a standard mask, not the general oscilloscope autoset
function.
Group
Mask
Syntax 1
MASK:AUTOSet:AUTOAdjust {ON|OFF|<NR1>}
Syntax 2
MASK:AUTOSet:AUTOAdjust?
Arguments
H ON
This enables the autoset:autoadjust function.
H OFF
This disables the autoset:autoadjust function.
H <NR1>
A 0 disables the autoset:autoadjust function; any other value enables
it.
Example
MASK:AUTOSet:AUTOAdjust ON
Enables the autoset autoadjust function

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish