EasyManuals Logo

Tektronix TDS5000B Series Online Programmer Guide

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #598 background imageLoading...
Page #598 background image
TDS5000B Online Programmer Guide
-- 5 7 8 --
MEASUrement:MEAS<x>:REFLevel:PERCent:LOW
Description
This command sets or queries the percentage (where 100% is equal to
HIGH) used to calculate the low reference level when MEASUre-
ment:MEAS<x>:REFLevel:METHod is set to Percent. This command
is equivalent to selecting Reference Levels from the Measure menu,
and then entering the Percentage Low Ref value. Measurements are
specified by x, which ranges from 1 through 8.
Note: This command affects the results of rise and fall measurements.
Group
Measurement
Related Commands
MEASUrement:MEAS<x>:REFLevel:METHod
Syntax 1
MEASUrement:MEAS<x>:REFLevel:PERCent:LOW <NR3>
Syntax 2
MEASUrement:MEAS<x>:REFLevel:PERCent:LOW?
Arguments
H <NR3>
This is the low reference level, ranging from 0 to 100%. The default
low reference level is 10%.
Example 1
MEASUREMENT:MEAS<x>:REFLEVEL:PERCENT:LOW 15
This command sets the high reference level to 15% of HIGH.
Example 2
MEASUREMENT:MEAS<x>:REFLEVEL:PERCENT:LOW?
This query might return
:MEASUREMENT:MEAS1:REFLEVEL:PERCENT:LOW 10
indicating that the percentage high reference level is set to 10% of
HIGH.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals