Page #271 background imageLoading...
Page #271 background image
TDS5000B Online Programmer Guide
-- 2 5 1 --
DISplay:SCREENTExt:LABel<x>:YPOS
Description
This command sets the vertical position of a given label. The label is
specified by x. The value of x can range from 1 through 8.
Group
Display Control
Related Commands
DISplay:SCREENTExt?, DISplay:SCREENTExt:STATE, DIS-
play:SCREENTExt:LABel<x>?,
DISplay:SCREENTExt:LABel<x>:NAMe, DISplay:SCREEN-
TExt:LABel<x>:XPOS
Syntax
<NR1>
Argument
H <NR1>
This argument sets the vertical position of the screen text label. The
value can range from 0 to 385.
Example 1
DISplay:SCREENTExt:LABel3:YPOS 100
This command sets the vertical position of label 3 to 100.
Example 2
DISPLAY:SCREENTEXT:LABEL3:YPOS?
This query might return
:DISPLAY:SCREENTEXT:LABEL3:YPOS 192
indicating that screen label 3 is vertically positioned to 192 (midscreen
vertically).

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Tektronix TDS5000B Series Specifications

General IconGeneral
Channels2 or 4
Vertical Resolution8 bits
Display TypeColor LCD
Vertical Sensitivity1 mV/div to 10 V/div
Input CouplingAC, DC, GND
Timebase Accuracy± 50 ppm
Operating Temperature0 °C to +50 °C
Bandwidthup to 1 GHz
Sample Rateup to 5 GS/s
Display Size10.4 inches
Input Impedance1 MΩ ± 1%
Trigger TypesEdge, Pulse Width, Logic, Runt, Timeout, Setup/Hold
Maximum Input Voltage300 V (DC + peak AC)
InterfaceUSB, GPIB, Ethernet