TDS5000B Online Programmer Guide
-- 7 7 8 --
TRIGger:B:BY
Description
This command selects or returns whether the B trigger occurs after a
specified number of events or a specified period of time after the A
trigger. This is equivalent to selecting B Event (Delayed) Trigger Setup
from the Trig menu, selecting the A→B Seq tab, and then choosing
Trig After Time or Trig on nth event.
Note: The traditional Runs After functionality is now served by the
Horizontal Delay function. For details, see the HORIZon-
tal[:MAIN]:DELay:MODe and HORIZontal[:MAIN]:DELay:TIMe
commands.
Group
Trigger
Related Commands
TRIGger:B:EVENTS:COUNt, TRIGger:B:TIMe, HORIZon-
tal[:MAIN]:DELay:MODe, HORIZontal[:MAIN]:DELay:TIMe
Syntax 1
TRIGger:B:BY {EVENTS|TIMe}
Syntax 2
TRIGger:B:BY?
Arguments
H EVENTS
This sets the B trigger to take place following a set number of trigger
events after the A trigger occurs. The number of events is specified by
TRIGger:B:EVENTS:COUNt.
H TIMe
This sets the B trigger to occur a set time after the A trigger event.
The time period is specified by TRIGger:B:TIMe.
Example 1
TRIGger:B:BY?
This query might return
:TRIGGER:B:BY EVENTS
indicating that the B trigger takes place following a set number of trig-
ger events after the A trigger occurs.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals