MEASUrement:MEAS<x>:REFLevel:PERCent:HIGH
Description
This command sets or queries the percentage (where 100% is equal to
HIGH) used to calculate the high reference level when MEASUre-
ment:MEAS<x>:REFLevel:METHod is set to Percent. Measurements
are specified by x, which ranges from 1 through 8.
Note: This command affects the results of rise and fall measurements.
Group
Measurement
Related Commands
MEASUrement:MEAS<x>:REFLevel:METHod
Syntax 1
MEASUrement:MEAS<x>:REFLevel:PERCent:HIGH
<NR3>
Syntax 2
MEASUrement:MEAS<x>:REFLevel:PERCent:HIGH?
Arguments
H <NR3>
This is the high reference level, ranging from 0 to 100%. The default
high reference level is 90%.
Example 1
MEASUREMENT:MEAS1:REFLEVEL:PERCENT:HIGH 95
This command sets the high reference level to 95% of HIGH.
Example 2
MEASUREMENT:MEAS1:REFLEVEL:PERCENT:HIGH?
This query might return
:MEASUREMENT:MEAS1:REFLEVEL:PERCENT:HIGH 90
indicating that the percentage high reference level is set to 90% of
HIGH.