MATH<x>:SPECTral:CENTER
Description
This command specifies or returns the center frequency of the spectral
analyzer output data span for the specified math waveform. The Math
waveform is specified by x, which ranges from 1 through 4 for four--
channel instruments or 1 through 2 for two--channel instruments. This
command is equivalent to selecting Spectral Setup from the Math menu
and then entering a Center Freq value.
Group
Math
Related Commands
MATH<x>:SPECTral:SPAN
Syntax 1
MATH<x>:SPECTral:CENTER <NR3>
Syntax 2
MATH<x>:SPECTral:CENTER?
Argument
H <NR3>
This is the desired frequency of the spectral analyzer output data span
in hertz.
Example 1
MATH3:SPECTral:CENTER 10.09E6
This command sets the center frequency to the closest value it can at-
tain to 10.09 MHz.
Example 2
MATH2:SPECTral:CENTER?
This query might return
:MATH2:SPECTral:CENTER 10.0900E+06
indicating that the center frequency is currently set at 10.09 MHz.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Tektronix TDS5000B Series Specifications

General IconGeneral
Channels2 or 4
Vertical Resolution8 bits
Display TypeColor LCD
Vertical Sensitivity1 mV/div to 10 V/div
Input CouplingAC, DC, GND
Timebase Accuracy± 50 ppm
Operating Temperature0 °C to +50 °C
Bandwidthup to 1 GHz
Sample Rateup to 5 GS/s
Display Size10.4 inches
Input Impedance1 MΩ ± 1%
Trigger TypesEdge, Pulse Width, Logic, Runt, Timeout, Setup/Hold
Maximum Input Voltage300 V (DC + peak AC)
InterfaceUSB, GPIB, Ethernet

Related product manuals