EasyManuals Logo

Tektronix TDS5000B Series Online Programmer Guide

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #581 background imageLoading...
Page #581 background image
TDS5000B Online Programmer Guide
-- 5 6 1 --
MEASUrement:IMMed:VALue?
Description
This query--only command returns the value of the measurement speci-
fied by the MEASUrement:IMMed:TYPe command. The measurement
is immediately taken on the source(s) specified by a MEASUre-
ment:IMMed:SOURCE command.
Note: A change to HORizontal:MAIn:SCALe or CH<x>:SCALe will
not necessarily have taken affect if followed by this command.
Group
Measurement
Related Commands
MEASUrement:IMMed:TYPe, MEASUrement:IMMed:SOURCE,
*ESR?, ALLEV?
Syntax
MEASUrement:IMMed:VALue?
Example
MEASUrement:IMMed:VALue?
This query might return
:MEASUREMENT:IMMED:VALUE 9.9000E+37
If the measurement has an error or warning associated with it, then an
item is added to the error queue. The error can be checked for with the
*ESR? and ALLEV? commands.