TDS5000B Online Programmer Guide
-- 3 8 8 --
LIMit:LOCk
Description
This command sets or queries whether vertical scaling and positioning
affect both source and template for template comparison pairs.
Group
Limit Test
Related Commands
Syntax 1
LIMit:LOCk {OFF|ON|<NR1>}
Syntax 2
LIMit:LOCk?
Arguments
H OFF
This argument disables the lock feature.
H ON
This argument enables the lock feature.
H NR1
A 0 disables the lock feature; any other value enables the lock feature.
For queries, a 0 is returned if the lock feature is off; a 1 is returned if
the lock feature is on.
Example 1
LIMit:LOCk ON
This command enables the lock feature so that vertical scaling and posi-
tioning affect both the source and the template for template comparison
pairs.
Example 2
LIMit:LOCk?
This query might return:
:LIMIT:LOCK 0
This indicates that the lock feature is disabled.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Tektronix TDS5000B Series Specifications

General IconGeneral
Channels2 or 4
Vertical Resolution8 bits
Display TypeColor LCD
Vertical Sensitivity1 mV/div to 10 V/div
Input CouplingAC, DC, GND
Timebase Accuracy± 50 ppm
Operating Temperature0 °C to +50 °C
Bandwidthup to 1 GHz
Sample Rateup to 5 GS/s
Display Size10.4 inches
Input Impedance1 MΩ ± 1%
Trigger TypesEdge, Pulse Width, Logic, Runt, Timeout, Setup/Hold
Maximum Input Voltage300 V (DC + peak AC)
InterfaceUSB, GPIB, Ethernet