TDS5000B Online Programmer Guide
-- 811 --
WFMInpre:YUNit
Description
This command sets or queries the vertical units of the incoming wave-
form.
Group
Waveform Transfer
Related Commands
WFMOutpre:YUNit
Syntax 1
WFMInpre:YUNit <String>
Syntax 2
WFMInpre:YUNit?
Argument
H <String>
This contains a maximum of three alpha characters that represent the
vertical unit of measure for the incoming waveform.
Example 1
WFMInpre:YUNit?
This query might return
:WFMINPRE:YUNIT ”s”
indicating the vertical units for the incoming waveform are seconds.
Example 2
WFMInpre:YUNit ”Pa”
This command specifies that the vertical units for the incoming wave-
form are Pascal.