-- 6 7 2 --
TRIGger:A:EDGE?
Description
This query--only command returns the trigger source, coupling, and
slope for the A edge trigger. This command is equivalent to selecting
Edge Setup from the Trig menu and viewing the current setups.
Group
Trigger
Related Commands
TRIGger:A:PULse, TRIGger:A:LOGIc
Syntax
TRIGger:A:EDGE?
Example
TRIGger:A:EDGE?
This query might return
:TRIGGER:A:EDGE:SOURCE CH1;COUPLING DC;
SLOPE RISE
indicating the trigger source, coupling, and slope for the A edge trigger.