EasyManuals Logo
Home>Tektronix>Test Equipment>TDS5000B Series

Tektronix TDS5000B Series Online Programmer Guide

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #708 background imageLoading...
Page #708 background image
TDS5000B Online Programmer Guide
-- 6 8 8 --
TRIGger:A:LOGIc:PATtern:WHEn:LESSLimit
Description
This command sets or queries the maximum time that the selected pat-
tern may be true and still generate an A logic pattern trigger. This com-
mand is equivalent to selecting A Event (Main) Trigger Setup from the
Trig menu, selecting Pattern as the Trigger Type, selecting Less Than
for the Pattern in the Trigger When settings, and entering a maximum
value for Time.
Group
Trigger
Related Commands
TRIGger:A:LOGIc:PATtern:WHEn:MORELimit
Syntax 1
TRIGger:A:LOGIc:PATtern:WHEn:LESSLimit <NR3>
Syntax 2
TRIGger:A:LOGIc:PATtern:WHEn:LESSLimit?
Argument
H <NR3>
This specifies the amount of time to hold the pattern true.
Example 1
TRIGger:A:LOGIc:PATtern:WHEn:LESSLimit?
This query might return
:TRIGGER:A:LOGIC:PATTERN:WHEN:LESSLIMIT
5.0000E-09
indicating that the selected pattern may hold true for up to 5 ns and still
generate an A logic pattern trigger.
Example 2
TRIGger:A:LOGIc:PATtern:WHEn:LESSLimit
10.0E+00
This command sets the maximum time that the selected pattern may
hold true (and generate an A logic pattern trigger) to 10 s.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals