TDS5000B Online Programmer Guide
-- 4 9 4 --
MASK:USER:TRIGTOSAMP
Description
This command sets or returns the time, in seconds, from the (leading
edge) trigger position to the pulse bit sampling position. The query
form of this command returns the time value of the displayed mask.
The set form of this command only affects the User mask, regardless of
the current (displayed) mask.
Group
Mask
Related Commands
MASK:USER:WIDth
Syntax 1
MASK:USER:TRIGTOSAMP <NR3>
Syntax 2
MASK:USER:TRIGTOSAMP?
Argument
H <NR3>
Is the floating point number that sets the time to the pulse bit sam-
pling position.
Example
MASK:USER:TRIGTOSAMP 162e-9
Sets the User mask time from the (leading edge) trigger position to the
pulse bit sampling position to 162 nanoseconds.