TDS5000B Online Programmer Guide
-- 5 4 3 --
MEASUrement:IMMed:REFLevel:PERCent:LOW
Description
This command sets or queries the percentage (where 100% is equal to
HIGH) used to calculate the low reference level when MEASUre-
ment:IMMed:REFLevel:METHod is set to Percent. Note that this com-
mand affects the results of rise and fall measurements. This command is
equivalent to selecting Reference Levels from the Measure menu and
then entering the Percentage Low Ref value.
Group
Measurement
Related Commands
MEASUrement:IMMed:REFLevel:METHod
Syntax 1
MEASUrement:IMMed:REFLevel:PERCent:LOW <NR1>
Syntax 2
MEASUrement:IMMed:REFLevel:PERCent:LOW?
Arguments
H <NR1>
This is the low reference level, ranging from 0 to 100%. The default
low reference level is 10%.
Example 1
MEASUREMENT:IMMED:REFLEVEL:PERCENT:LOW 15
This command sets the low reference level to 15% of HIGH.
Example 2
MEASUREMENT:IMMED:REFLEVEL:PERCENT:LOW?
This query might return
:MEASUREMENT:IMMED:REFLEVEL:PERCENT:LOW 10
indicating that the percentage low reference level is set to 10% of
HIGH.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Tektronix TDS5000B Series Specifications

General IconGeneral
Channels2 or 4
Vertical Resolution8 bits
Display TypeColor LCD
Vertical Sensitivity1 mV/div to 10 V/div
Input CouplingAC, DC, GND
Timebase Accuracy± 50 ppm
Operating Temperature0 °C to +50 °C
Bandwidthup to 1 GHz
Sample Rateup to 5 GS/s
Display Size10.4 inches
Input Impedance1 MΩ ± 1%
Trigger TypesEdge, Pulse Width, Logic, Runt, Timeout, Setup/Hold
Maximum Input Voltage300 V (DC + peak AC)
InterfaceUSB, GPIB, Ethernet

Related product manuals