TDS5000B Online Programmer Guide
-- 7 2 --
*SDS
Description
This command (no query form) changes the specified setup to reference
the factory setup instead of the specific user setup slot. The content of
the setup slot is unchanged, but the data will no longer be accessible to
you.
This command is equivalent to selecting Delete from the File menu,
and then choosing the All Setups.
Group
Save and Recall
Related Commands
DELEte:SETUp
Syntax
*SDS <NR1>
Argument
H <NR1>
This specifies a user setup location to delete. Setup storage location
values range from 1 through 10; using an out--of--range value causes
an error.
Example
*SDS 2
This command changes setup slot 2 to reference the factory setup.