TDS5000B Online Programmer Guide
-- 7 0 0 --
TRIGger:A:LOGIc:SETHold?
Description
This query--only command returns the clock edge polarity, voltage
threshold and source input; data voltage threshold and source; and both
setup and hold times for setup and hold violation triggering. This com-
mand is equivalent to selecting Setup/Hold Setup from the Trig menu
and then viewing the current setups.
Group
Trigger
Related Commands
TRIGger:A:LOGIc:PATtern?, TRIGger:A:LOGIc:STATE?
Syntax
TRIGger:A:LOGIc:SETHold?
Example
TRIGger:A:LOGIc:SETHold?
This query might return the settings
:TRIGGER:A:LOGIC:SETHOLD:CLOCK:EDGE RISE;
THRESHOLD 1.4000;SOURCE CH2;
:TRIGGER:A:LOGIC:SETHOLD:DATA:THRESHOLD
1.4000;SOURCE CH1;:TRIGGER:A:LOGIC:SETHOLD
:HOLDTIME 2.0000E-9;SETTIME 3.0000E-9