TDS5000B Online Programmer Guide
-- 5 5 3 --
MEASUrement:IMMed:TYPe
Description
This command sets or queries the immediate measurement type.
Group
Measurement
Syntax 1
MEASUrement:IMMed:TYPE {AMPlitude|AREa|
BURst|CARea|CMEan|CRMs|DELay|DISTDUty|
EXTINCTDB|EXTINCTPCT|EXTINCTRATIO|
EYEHeight|EYEWidth|FALL|FREQuency|HIGH|
HITs|LOW|MAXimum|MEAN|MEDian|MINImum|
NCROss|NDUty|NOVershoot|NWIdth|PBASe|
PCROss|PCTCROss|PDUty|PEAKHits|PERIod|
PHAse|PK2Pk|PKPKJitter|PKPKNoise|
POVershoot|PTOT|PWIdth|QFACtor|RISe|
RMS|RMSJitter|PMSNoise|SIGMA1|SIGMA2|
SIGMA3|SIXSigmajit|SNRatio|STDdev|
UNDEFINED|WAVEFORMS}
Syntax 2
MEASUrement:IMMed:TYPe?
Arguments
H AMPlitude
This measures the amplitude of the selected waveform. In other
words, it measures the high value less the low value measured over
the entire waveform or gated region.
Amplitude = High -- Low
H AREa
This measures the voltage over time. The area is over the entire wave-
form or gated region and is measured in volt--seconds. The area mea-
sured above the ground is positive, while the area below ground is
negative.
H BURst
The burst width measures the duration of a burst. The measurement is
made over the entire waveform or gated region.
H CARea
The cycle area measures the voltage over time. In other words, it mea-
sures, in volt--seconds, the area over the first cycle in the waveform or
the first cycle in the gated region. The area measured above the com-
mon reference point is positive, while the area below the common
reference point is negative.
H CMEan
This measures the arithmetic mean over the first cycle in the wave-
form or the first cycle in the gated region.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Related product manuals