TDS5000B Online Programmer Guide
-- 7 5 3 --
TRIGger:A:PULse:WINdow:THReshold?
Description
This query--only command returns the upper and lower thresholds for
the window trigger. This command query is equivalent to selecting
Window Setup from the Trig menu and viewing the window trigger
Upper Level and Lower Level settings.
Group
Trigger
Syntax
TRIGger:A:PULse:WINdow:THReshold?
Example
TRIGger:A:PULse:WINdow:THReshold?
This query might return
:TRIGGER:A:PULSE:WINDOW:THRESHOLD:HIGH
1.2000;LOW 800.0000E-3
indicating that the upper threshold is 1.2 V and that the lower threshold
is 800 mV.