TDS5000B Online Programmer Guide
-- 7 1 7 --
TRIGger:A:PULse:RUNT:LOGIc:THReshold:CH<x>
Description
This command sets or queries the A runt logic threshold for the channel
specified by <x>, which can be 1, 2, 3, or 4. This is equivalent to select-
ing Runt Setup from the Trig menu and setting the runt logic threshold.
While all channels can be set or queried, only channels 3 and 4 can be
used if the runt source is channel 1 or 2. Similarly only channels 1 and
2 can be used if the runt source is channel 3 or 4.
Note: This command is not available in two--channel instruments.
Group
Trigger
Syntax 1
TRIGger:A:PULse:RUNT:LOGIc:THReshold:CH<x>
<NR3>
Syntax 2
TRIGger:A:PULse:RUNT:LOGIc:THReshold:CH<x>?
Argument
H <NR3>
This argument specifies the threshold in volts.
Example 1
TRIGger:A:PULse:RUNT:LOGIc:THReshold:CH1?
This query might return
:TRIGGER:A:PULSE:RUNT:LOGIC:THRESHOLD:CH1
0.5000
indicating that the runt logic trigger threshold is set to 0.5 V for chan-
nel 1.
Example 2
TRIGger:A:PULse:RUNT:LOGIc:THReshold:CH3 0.5.
This command sets the runt logic trigger threshold to 0.5 V for chan-
nel 3.