-- 5 3 9 --
MEASUrement:IMMed:REFLevel:ABSolute:MID[1]
Description
This command sets or queries the mid reference level, and is the 50%
reference level when MEASUrement:IMMed:REFLevel:METHod is
set to Absolute. Note that this command affects the results of period,
frequency, delay, and all cyclic measurements.
Group
Measurement
Related Commands
MEASUrement:IMMed:REFLevel:METHod
Syntax 1
MEASUrement:IMMed:REFLevel:ABSolute:MID[1]
<NR3>
Syntax 2
MEASUrement:IMMed:REFLevel:ABSolute:MID[1]?
Arguments
H <NR3>
This is the mid reference level, in volts. The default is 0.0 V.
Example 1
MEASUREMENT:IMMED:REFLEVEL:ABSOLUTE:MID[1] .71
This command sets the mid reference level to .71 V.
Example 2
MEASUREMENT:IMMED:REFLEVEL:ABSOLUTE:MID[1]?
This query might return
:MEASUREMENT:IMMED:REFLEVEL:ABSOLUTE:MID
0.7100E+00
indicating that the absolute mid1 reference level is set to .71 V.