EasyManuals Logo

Tektronix TDS5000B Series Online Programmer Guide

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #715 background imageLoading...
Page #715 background image
TDS5000B Online Programmer Guide
-- 6 9 5 --
TRIGger:A:LOGIc:SETHold:DATa:SOUrce
Description
This command sets or queries the data source for the setup and hold
trigger. This command is equivalent to selecting Setup/Hold Setup from
the Trig menu and choosing the desired channel from the Data Source
drop--down list.
Group
Trigger
Related Commands
TRIGger:A:LOGIc:SETHold:DATa:THReshold
Syntax 1
TRIGger:A:LOGIc:SETHold:DATa:SOUrce CH<x>
Syntax 2
TRIGger:A:LOGIc:SETHold:DATa:SOUrce?
Argument
H CH<x>
This specifies the input channel, which ranges from 1 through 4 for
four--channel instruments or 1 through 2 for two--channel instru-
ments..
Example 1
TRIGger:A:LOGIc:SETHold:DATa:SOUrce?
This query might return
:TRIGGER:A:LOGIC:SETHOLD:DATA:SOURCE CH2
indicating that channel 2 is the current clock source for the setup and
hold trigger.
Example 2
TRIGger:A:LOGIc:SETHold:DATa:SOUrce CH1
This command sets channel 1 as the clock source for the setup and hold
trigger.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals