TDS5000B Online Programmer Guide
-- 4 8 1 --
MASK:TESt:THReshold
Description
This command sets or returns the number of failed tested waveforms
needed in each pass/fail test to cause the pass/fail status to change from
PASSING to FAILING.
Group
Mask
Syntax 1
MASK:TESt:THReshold <NR1>
Syntax 2
MASK:TESt:THReshold?
Argument
H <NR1>
Is an integer that represents the number of tested waveform violations
occurring in each pass/fail test needed to change the pass/fail status to
FAILING.
This means that if the threshold is set to the default of 1 then 1 failure
would cause a test failure. If the threshold is set to 2 then 2 failures
would cause a test failure.
Example
MASK:TEST:THRESHOLD?
Might return: MASK:TEST:THRESHOLD 32