TDS5000B Online Programmer Guide
-- 2 1 3 --
DIAg:SELect:LAST
Description
This command (no query form) selects one or more diagnostic menu
items to be executed via the DIAg:EXECUTE command. If you specify
DIAg:LEVEL SUBSYS, then menu items come from this diagnostic
level and are limited to the value returned by the DIAg:NUMITEMS?
query. For example, specifying DIAg:SELECT:SUBSYS 2 and
DIAg:NUMITEMS 5 indicates that diagnostics will start from subsys-
tem 2 and that you can specify a range from 2 through 5 for DIAg:SE-
Lect:LAST.
If you enter:
DIAg:SELect:LAST 2, only subsystem 2 will be executed.
DIAg:SELect:LAST 4, subsystems 2 through 4 will be executed.
Group
Diagnostics
Related Commands
DIAg:EXECUTE, DIAg:LEVEL, DIAg:NAMe:AREA?,
DIAg:NAMe:SUBSYS?, DIAg:NAMe:TEST?, DIAg:NUMITEMS?,
DIAg:SELect:SUBSYS
Syntax
DIAg:SELect:LAST <NR1>
Arguments
H <NR1>
This selects an integer that identifies the number of the last item that
will be executed when the DIAg:EXECUTE command is run.
Example
DIAg:SELect:LAST 2
This command specifies that (based on the previous example) only sub-
system 2 will be executed.