EasyManua.ls Logo

Tektronix TDS5000B Series Online Programmer Guide

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #102 background imageLoading...
Page #102 background image
TDS5000B Online Programmer Guide
-- 8 2 --
ACQuire:NUMEnv
Description
This command sets or queries the number of waveform acquisitions
that make up an envelope waveform. Sending this command is equiva-
lent to setting the # of Wfms in the Acquisition Setup menu when En-
velope Acquisition mode is selected.
Group
Acquisition
Syntax 1
ACQuire:NUMEnv {<NR1>|INFInite}
Syntax 2
ACQuire:NUMEnv?
Arguments
H NR1
This is the number of waveform acquisitions to be enveloped.
Example 1
ACQuire:NUMenv 10
This command specifies that an enveloped waveform will show the
result of combining 10 separately acquired waveforms.
Example 2
ACQuire:NUMenv?
This command might return
:ACQUIRE:NUMENV 0
indicating that acquisitions are acquired infinitely for enveloped wave-
forms.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Tektronix TDS5000B Series Specifications

General IconGeneral
Channels2 or 4
Vertical Resolution8 bits
Display TypeColor LCD
Vertical Sensitivity1 mV/div to 10 V/div
Input CouplingAC, DC, GND
Timebase Accuracy± 50 ppm
Operating Temperature0 °C to +50 °C
Bandwidthup to 1 GHz
Sample Rateup to 5 GS/s
Display Size10.4 inches
Input Impedance1 MΩ ± 1%
Trigger TypesEdge, Pulse Width, Logic, Runt, Timeout, Setup/Hold
Maximum Input Voltage300 V (DC + peak AC)
InterfaceUSB, GPIB, Ethernet