EasyManua.ls Logo

Tektronix TDS5000B Series

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TDS5000B Online Programmer Guide
-- 5 3 7 --
MEASUrement:IMMed:REFLevel:ABSolute:HIGH
Description
This command sets or queries the high reference level, and is the upper
reference level when MEASUrement:IMMed:REFLevel:METHod is
set to Absolute. Note that this command affects the results of rise and
fall measurements.
Group
Measurement
Related Commands
MEASUrement:IMMed:REFLevel:METHod
Syntax 1
MEASUrement:IMMed:REFLevel:ABSolute:HIGH <NR3>
Syntax 2
MEASUrement:IMMed:REFLevel:ABSolute:HIGH?
Arguments
H <NRf>
This is the high reference level, in volts. The default is 0.0 V.
Example 1
MEASUREMENT:IMMed:REFLEVEL:ABSOLUTE:HIGH 1.71
This command sets the high reference level to 1.71 V.
Example 2
MEASUREMENT:IMMED:REFLEVEL:ABSOLUTE:HIGH?
This query might return
:MEASUREMENT:IMMED:REFLEVEL:ABSOLUTE:HIGH
1.7100E+00
indicating that the absolute high reference level is set to 1.71 V.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Related product manuals