TDS5000B Online Programmer Guide
-- 8 2 6 --
WFMOutpre:XINcr?
Description
This query--only command returns the horizontal point spacing in units
of WFMOutpre:XUNit for the waveform specified by the
DATa:SOUrce command. This value corresponds to the sampling inter-
val.
An error is reported if the DATa:SOUrce waveform does not exist.
Group
Waveform Transfer
Related Commands
DATa:SOUrce, WFMOutpre:XUNit,
Syntax
WFMOutpre:XINcr?
Example
WFMOutpre:XINcr?
This query might return
:WFMOUTPRE:XINCR 10.0000E-6
indicating that the horizontal sampling interval is 10 µs/point (500
µs/div).