CALibrate:RESults?
Description
This query--only command returns the status of internal and factory
calibrations, without performing any calibration operations. The results
returned do not include the calibration status of attached probes. The
CALibration:RESults? query is intended to support GO/NoGO testing
of the oscilloscope calibration readiness: all returned results should in-
dicate PASS status if the instrument is ”fit for duty”. It is quite com-
mon, however, to use uncalibrated probes (particularly when the instru-
mentt inputs are connected into a test system with coaxial cables).
This command is equivalent to selecting Instrument Calibration from
the Utilities menu and then viewing the contents of the Status field.
Group
Calibration
Related Commands
*CAL?
Syntax
CALibrate:RESults?
Example
CALibrate:RESults?
This query returns the status of all the calibration subsystems. The
query might return :
:CALIBRATE:RESULTS:SPC PASS;FACTORY PASS;FAC-
TORY:VOLTAGE PASS;FREQUENCY PASS;TRIGGER PASS

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals