EasyManua.ls Logo

Tektronix TDS5000B Series

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TDS5000B Online Programmer Guide
-- 7 3 8 --
TRIGger:A:PULse:TRANsition:WHEn
Description
This command sets or queries whether to check for a transitioning sig-
nal that is faster or slower than the specified delta time. This is equiva-
lent to selecting Transition Setup from the Trig menu and choosing the
Trigger When Transition Time setting.
Group
Trigger
Related Commands
TRIGger:A:PULse:TRANsition:DELTATime, TRIG-
ger:A:PULse:TRANsition:POLarity, TRIGger:A:PULse:TRANsi-
tion:THReshold?
Syntax 1
TRIGger:A:PULse:TRANsition:WHEn {FASTERt-
han|SLOWERthan}
Syntax 2
TRIGger:A:PULse:TRANsition:WHEn?
Arguments
H FASTERthan
This sets the trigger to occur when the transitioning signal is faster
than the set volts/second rate.
H SLOWERthan
This sets the trigger to occur when the transitioning signal is slower
than the set volts/second rate.
Example 1
TRIGger:A:PULse:TRANsition:WHEn?
This query might return
:TRIGGER:A:PULSE:TRANSITION:WHEN FASTERTHAN
indicating that the transition triggers when the transitioning signal is
faster than the set volts/second rate.
Example 2
TRIGger:A:PULse:TRANsition:WHEn SLOWERthan

Table of Contents