TDS5000B Online Programmer Guide
-- 8 0 4 --
WFMInpre:PT_Off
Description
This command specifies or returns the trigger point within the wave-
form record for the reference waveform specified by the DATa:DES-
Tination command.
Group
Waveform Transfer
Related Commands
DATa:DESTination, DATa:STARt, WFMOutpre:PT_Off
Syntax 1
WFMInpre:PT_Off <NR1>
Syntax 2
WFMInpre:PT_Off?
Argument
H <NR1>
This is (record length --1) to record length and is the position of the
data point immediately following the actual trigger. <NR1> is ex-
pressed relative to DATa:STARt.
Example 1
WFMInpre:PT_Off 0
This command specifies that the trigger point is the first point in the
waveform record, which is the default.
Example 2
WFMInpre:PT_Off?
This query might return
:WFMINPRE:PT_OFF 0
indicating that the incoming waveform trigger point is the first point in
the waveform record.