-- 4 3 5 --
MASK:MARgin:PERCent
Description
This command sets or queries the mask margin percentage. If you set a
positive mask margin percentage, the polygon expands and the mask
test will be more difficult to pass due to the expanded test area. If you
set a negative mask margin percentage, the polygon shrinks and the
mask test will be easier to pass due to the smaller area.
Group
Mask
Syntax 1
MASK:MARgin:PERCent <NR3>
Syntax 2
MASK:MARgin:PERCent?
Argument
H <NR3>
This values range from –50.0 to +50.0. The default is 5.
Example 1
MASK:MARGIN:PERCENT 25
Sets the mask margin percentage to 25.
Example 2
MASK:MARGIN:PERCENT?
Might return
:MASK:MARGIN:PERCENT 1.00000000000E+001,
indicating that the mask margin is set to 10%.

Table of Contents