-- 7 5 5 --
TRIGger:A:PULse:WINdow:WHEn
Description
This command sets or queries the window trigger violation qualifica-
tion. This command is equivalent to selecting Window Setup from the
Trig menu and selecting Logic, Occurs, or Wider than in the Trigger
When box.
Group
Trigger
Related Commands
Syntax 1
TRIGger:A:PULseWINdow:WHEn {LOGIc|OCCurs|WID-
ERthan}
Syntax 2
TRIGger:A:PULse:WINdow:WHEn?
Arguments
H LOGIc
This argument specifies a trigger event when a window violation oc-
curs on the AND of the logic channels.
H OCCurs
This argument specifies a trigger event if any detectable window
violation occurs.
H WIDERthan
This specifies a trigger event if a window violation greater than the
specified width occurs.
Example 1
TRIGger:A:PULse:WINdow:WHEn?
This query might return
:TRIGGER:A:PULSE:WINdow:WHEN OCCURS
indicating that a trigger will occur if the instrument detects any window
violation.
Example 2
TRIGger:A:PULse:WINdow:WHEn OCCurs
This command sets the widow violation qualifier to OCCurs, which
causes a trigger to occur when the instrument detects any window
violation.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals