EasyManua.ls Logo

Tektronix TDS5000B Series Online Programmer Guide

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #776 background imageLoading...
Page #776 background image
TDS5000B Online Programmer Guide
-- 7 5 6 --
TRIgger:A:PULse:WINdow:WIDTH
Description
This command sets or queries the minimum width for a window viola-
tion. This command is equivalent to selecting Window Setup from the
Trig menu, selecting Wider than in the Trigger When box, and setting
the Width.
Group
Trigger
Syntax 1
TRIGger:A:PULse:WINdow:WIDTH <NR3>
Syntax 2
TRIGger:A:PULse:WINdow:WIDTH?
Argument
H <NR3>
This argument specifies the minimum width in seconds.
Example 1
TRIGger:A:PULse:WINdow:WIDTH?
This query might return
:TRIGGER:A:PULSE:WINdow:WIDTH 2.0000E-09
indicating that the minimum width of a valid window violation is 2 ns.
Example 2
TRIGger:A:PULse:WINdow:WIDTH 15E-9
This command sets the minimum width of a valid window violation to
15 ns.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Tektronix TDS5000B Series Specifications

General IconGeneral
Channels2 or 4
Vertical Resolution8 bits
Display TypeColor LCD
Vertical Sensitivity1 mV/div to 10 V/div
Input CouplingAC, DC, GND
Timebase Accuracy± 50 ppm
Operating Temperature0 °C to +50 °C
Bandwidthup to 1 GHz
Sample Rateup to 5 GS/s
Display Size10.4 inches
Input Impedance1 MΩ ± 1%
Trigger TypesEdge, Pulse Width, Logic, Runt, Timeout, Setup/Hold
Maximum Input Voltage300 V (DC + peak AC)
InterfaceUSB, GPIB, Ethernet

Related product manuals