TDS5000B Online Programmer Guide
-- 7 5 6 --
TRIgger:A:PULse:WINdow:WIDTH
Description
This command sets or queries the minimum width for a window viola-
tion. This command is equivalent to selecting Window Setup from the
Trig menu, selecting Wider than in the Trigger When box, and setting
the Width.
Group
Trigger
Syntax 1
TRIGger:A:PULse:WINdow:WIDTH <NR3>
Syntax 2
TRIGger:A:PULse:WINdow:WIDTH?
Argument
H <NR3>
This argument specifies the minimum width in seconds.
Example 1
TRIGger:A:PULse:WINdow:WIDTH?
This query might return
:TRIGGER:A:PULSE:WINdow:WIDTH 2.0000E-09
indicating that the minimum width of a valid window violation is 2 ns.
Example 2
TRIGger:A:PULse:WINdow:WIDTH 15E-9
This command sets the minimum width of a valid window violation to
15 ns.