EasyManua.ls Logo

Tektronix TDS5000B Series

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TDS5000B Online Programmer Guide
-- 7 4 1 --
TRIGger:A:PULse:WIDth:LOWLimit
Description
This command sets or queries the lower limit for the width trigger. This
command is equivalent to selecting Width Setup from the Trig menu
and setting the pulse Lower Limit.
Group
Trigger
Related Commands
TRIGger:A:PULse:WIDth?
Syntax 1
TRIGger:A:PULse:WIDth:LOWLimit <NR3>
Syntax 2
TRIGger:A:PULse:WIDth:LOWLimit?
Argument
H <NR3>
This specifies the A pulse width trigger lower limit, in seconds.
Example 1
TRIGger:A:PULse:WIDth:LOWLimit?
This query might return
:TRIGGER:A:PULSE:WIDTH:LOWLIMIT 1.0000E-9
indicating that the lower limit of the width trigger is set to 1 ns.
Example 2
TRIGger:A:PULse:WIDth:LOWLimit 2.0E-6
This command sets the lower limit of the width trigger to 2 µs.