TDS5000B Online Programmer Guide
-- 7 3 3 --
TRIGger:A:PULse:TRANsition:POLarity
Description
This command sets or queries the polarity for the transition trigger. This
command is equivalent to selecting Transition Setup from the Trig
menu and choosing from the Polarity pull --down list.
Group
Trigger
Related Commands
TRIGger:A:PULse:TRANsition:DELTATime
Syntax 1
TRIGger:A:PULse:TRANsition:POLarity {POS-
ITIVe|NEGative|EITher}
Syntax 2
TRIGger:A:PULse:TRANsition:POLarity?
Arguments
H POSITIVe
This indicates that a pulse edge must traverse from the lower (most
negative) to higher (most positive) level for transition triggering to
occur.
H NEGative
This indicates that a pulse edge must traverse from the upper (most
positive) to lower (most negative) level for transition triggering to
occur.
H EITher
This indicates either positive or negative polarity.
Example 1
TRIGger:A:PULse:TRANsition:POLarity?
This query might return
:TRIGGER:A:PULSE:TRANSITION:POLARITY EITHER
indicating that the polarity can be either positive or negative.
Example 2
TRIGger:A:PULse:TRANsition:DeltaTime NEGative
This command sets the transition polarity to negative.