TDS5000B Online Programmer Guide
-- 5 3 8 --
MEASUrement:IMMed:REFLevel:ABSolute:LOW
Description
This command sets or queries the low reference level, and is the zero
percent level when MEASUrement:IMMed:REFLevel:METHod is set
to Absolute. Note that this command affects the results of rise and fall
measurements.
Related Commands
MEASUrement:IMMed:REFLevel:METHod
Group
Measurement
Syntax 1
MEASUrement:IMMed:REFLevel:ABSolute:LOW <NR3>
Syntax 2
MEASUrement:IMMed:REFLevel:ABSolute:LOW?
Arguments
H <NR3>
This is the low reference level, in volts. The default is 0.0 V.
Example 1
MEASUREMENT:IMMED:REFLEVEL:ABSolute:LOW 0.0 V.
This command sets the low reference level to 0.0 V.
Example 2
MEASUREMENT:IMMED:REFLEVEL:ABSolute:LOW?
This query might return
:MEASUREMENT:MEAS1:REFLEVEL:ABSOLUTE:LOW
0.0000E+00
indicating that the absolute low reference level is set to 0.0 V.