EasyManuals Logo
Home>Tektronix>Test Equipment>TDS5000B Series

Tektronix TDS5000B Series Online Programmer Guide

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #192 background imageLoading...
Page #192 background image
TDS5000B Online Programmer Guide
-- 1 7 2 --
CURSor:XY:RECTY<x>
Description
This command sets or queries the Y cursor position in rectangular coor-
dinates. The cursor is specified by x, which can be 1 or 2.
Note: This query is only valid when the CURSor:XY:READOUT is
set to RECTangular. If the readout is set to other than rectangular
coordinates, the IEEE representation of NAN (Not a Number) is re-
turned: 99.1000E+36.
Group
Cursor
Syntax 1
CURSor:XY:RECTY<x> {<NR3>}
Syntax 2
CURSor:XY:RECTY<x>?
Arguments
H <NR3>
This is the Y position of the specified cursor in rectangular coordi-
nates.
Example 1
CURSor:XY:RECTY1 2.5
This command sets the Y position of Cursor 1 to 2.5 V.
Example 2
CURSor:XY:RECTY2?
This query might return
:CURSOR:XY:RECTY2 4.8800
indicating that the position of Cursor 2 in rectangular coordinates is
4.88 V.