TDS5000B Online Programmer Guide
-- 7 1 --
*SAV
Description
This command (no query form) stores the state of the instrument to a
specified memory location. You can later use the *RCL command to
restore the instrument to this saved state. \
This command is equivalent to selecting Save As from the File menu,
and then pressing the Setup button.
Group
Save and Recall
Related Commands
*RCL, RECAll:SETUp, SAVe:SETUp
Syntax
*SAV <NR1>
Argument
H <NR1>
This specifies a location in which to save the state of the instrument.
Location values range from 1 through 10. Using an out--of--range
location value causes an execution error. Any settings that have been
stored previously at this location will be overwritten.
Example
*SAV 2
This command saves the current instrument state in memory location 2.