EasyManuals Logo
Home>Tektronix>Test Equipment>TDS5000B Series

Tektronix TDS5000B Series Online Programmer Guide

Tektronix TDS5000B Series
917 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #713 background imageLoading...
Page #713 background image
TDS5000B Online Programmer Guide
-- 6 9 3 --
TRIGger:A:LOGIc:SETHold:CLOCk:THReshold
Description
This command sets or queries the clock voltage threshold for the setup
and hold trigger. This command is equivalent to selecting Setup/Hold
Setup from the Trig menu and setting the desired Clock Level.
Group
Trigger
Related Commands
TRIGger:A:LOGIc:SETHold:CLOCk:EDGE, TRIGger:A:LOGIc:SE-
THold:CLOCk:SOUrce
Syntax 1
TRIGger:A:LOGIc:SETHold:CLOCk:THReshold
{ECL|TTL|<NR3>}
Syntax 2
TRIGger:A:LOGIc:SETHold:CLOCk:THReshold?
Arguments
H ECL
This specifies a preset ECL high level of -- 1.3 V.
H TTL
This specifies a preset TTL high level of 1.4 V.
H <NR3>
This is the clock level, in volts.
Example 1
TRIGger:A:LOGIc:SETHold:CLOCk:THReshold?
This query might return
:TRIGGER:A:LOGIC:SETHOLD:CLOCK:THRESHOLD
1.2000E+00
indicating that the clock threshold for the setup and hold trigger is
1.2 V.
Example 2
TRIGger:A:LOGIc:SETHold:CLOCk:THReshold TTL
This command specifies the preset TTL value of 1.4 V as the clock
threshold for the setup and hold trigger.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals