TDS5000B Online Programmer Guide
-- 6 9 3 --
TRIGger:A:LOGIc:SETHold:CLOCk:THReshold
Description
This command sets or queries the clock voltage threshold for the setup
and hold trigger. This command is equivalent to selecting Setup/Hold
Setup from the Trig menu and setting the desired Clock Level.
Group
Trigger
Related Commands
TRIGger:A:LOGIc:SETHold:CLOCk:EDGE, TRIGger:A:LOGIc:SE-
THold:CLOCk:SOUrce
Syntax 1
TRIGger:A:LOGIc:SETHold:CLOCk:THReshold
{ECL|TTL|<NR3>}
Syntax 2
TRIGger:A:LOGIc:SETHold:CLOCk:THReshold?
Arguments
H ECL
This specifies a preset ECL high level of -- 1.3 V.
H TTL
This specifies a preset TTL high level of 1.4 V.
H <NR3>
This is the clock level, in volts.
Example 1
TRIGger:A:LOGIc:SETHold:CLOCk:THReshold?
This query might return
:TRIGGER:A:LOGIC:SETHOLD:CLOCK:THRESHOLD
1.2000E+00
indicating that the clock threshold for the setup and hold trigger is
1.2 V.
Example 2
TRIGger:A:LOGIc:SETHold:CLOCk:THReshold TTL
This command specifies the preset TTL value of 1.4 V as the clock
threshold for the setup and hold trigger.

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Tektronix TDS5000B Series Specifications

General IconGeneral
Channels2 or 4
Vertical Resolution8 bits
Display TypeColor LCD
Vertical Sensitivity1 mV/div to 10 V/div
Input CouplingAC, DC, GND
Timebase Accuracy± 50 ppm
Operating Temperature0 °C to +50 °C
Bandwidthup to 1 GHz
Sample Rateup to 5 GS/s
Display Size10.4 inches
Input Impedance1 MΩ ± 1%
Trigger TypesEdge, Pulse Width, Logic, Runt, Timeout, Setup/Hold
Maximum Input Voltage300 V (DC + peak AC)
InterfaceUSB, GPIB, Ethernet